Zebrafish Retina Electrode Recommendation |
For Larval Zebrafish Retina (which is quite small, 10microns) it may not be possible to inject DNA buffer.  Accordingly, we suggest that electroporation is performed after the larval zebrafish eye is taken out. Once this is done and one intends… |
|
Zebrafish Retina |
For Larval Zebrafish Retina (which is quite small, 10microns) it may not be possible to inject DNA buffer.  Accordingly, we suggest that electroporation is performed after the larval zebrafish eye is taken out. Once this is done and one intends… |
|
In Ovo Electroporation |
Some clients, however, who have to contend with a large number of samples on a daily basis find it difficult to manipulate a pair of electrodes vertically. Such clients prefer to move the pair of electrodes horizontally. The CUY613 series was designed… |
|
Mouse and Rat Small Tissue Electrode Recommendation and Protocol |
For smaller tissue of less than 0.5mm, we do not recommend the CUY520 series. The reason for this is practical; most clients want to minimize the required volume of DNA as much as possible and do not want to waste DNA. For such applications we believe… |
|
Mouse and Rat Small Tissue Electrode Recommendation and Protocol |
For smaller tissue of less than 0.5mm, we do not recommend the CUY520 series. The reason for this is practical; most clients want to minimize the required volume of DNA as much as possible and do not want to waste DNA. For such applications we believe… |
|
Mouse and Rat Tissue and Brain Slice |
… |
|
Mouse and Rate Tissue and Brain Slice |
… |
|
Cultured Mouse Embryos and Large Tissue |
For smaller tissue of less than 0.5mm, we do not recommend the CUY520 series. The reason for this is practical; most clients want to minimize the required volume of DNA as much as possible and do not want to waste DNA. For such applications we believe… |
|
Technical Drawing CUY701P2E |
|
|
CUY21SC - Square Wave Electroporator |
Because in vivo impedance (electrical resistance) is generally less stable than in vitro impedance, an accurate measurement of the target sample resistance prior to the electroporation event is the most important critical success factor for accurate… |
|
CUY21 Publication List, Electrode Recommendations and Protocol Information |
… |
|
CUY21 Publication List, Electrode Recommendations and Protocol Information |
… |
|
CUY21 Publication List, Electrode Recommendations and Protocol Information |
… |
|
CUY21 Publication List, Electrode Recommendations and Protocol Information |
… |
|
CUY21 Publication List, Electrode Recommendations and Protocol Information |
… |
|
CUY21 Publication List, Electrode Recommendations and Protocol Information |
… |
|
CUY21 Publication List, Electrode Recommendations and Protocol Information |
… |
|
CUY21 Publication List, Electrode Recommendations and Protocol Information |
… |
|
CUY21 Publication List, Electrode Recommendations and Protocol Information |
… |
|
CUY21 Publication List, Electrode Recommendations and Protocol Information |
… |
|
CUY21 Publication List, Electrode Recommendations and Protocol Information |
… |
|
NEPA21_Retina_EP |
(B) The current was applied with the positive electrode contralateral to the injected eye. After prior injection of plasmid DNA into the subretinal space of the right eye, this arrangement electrophoresed the negatively-charged DNA toward the RPE layer… |
|
Electroporation Protocols - In Vitro, In Ovo, In Utero, In Vivo, Ex Vivo and New Culture |
… |
|
|
… |
|
Electroporation Protocols - In Vitro, In Ovo, In Utero, In Vivo, Ex Vivo and New Culture |
… |
|
Electroporation Protocols - In Vitro, In Ovo, In Utero, In Vivo, Ex Vivo and New Culture |
… |
|
Electroporation Protocols - In Vitro, In Ovo, In Utero, In Vivo, Ex Vivo and New Culture |
… |
|
Electroporation Protocols - In Vitro, In Ovo, In Utero, In Vivo, Ex Vivo and New Culture |
… |
|
NEPA21_Retina_EP |
(B) The current was applied with the positive electrode contralateral to the injected eye. After prior injection of plasmid DNA into the subretinal space of the right eye, this arrangement electrophoresed the negatively-charged DNA toward the RPE layer… |
|
Electroporation Protocols - In Vitro, In Ovo, In Utero, In Vivo, Ex Vivo and New Culture |
… |
|
CUY21 Publication List, Electrode Recommendations and Protocol Information |
… |
|
CUY21 Publication List, Electrode Recommendations and Protocol Information |
… |
|
CUY21 Publication List, Electrode Recommendations and Protocol Information |
… |
|
CUY21 Publication List, Electrode Recommendations and Protocol Information |
… |
|
CUY21 Publication List, Electrode Recommendations and Protocol Information |
… |
|
CUY21 Publication List, Electrode Recommendations and Protocol Information |
… |
|
CUY21 Publication List, Electrode Recommendations and Protocol Information |
… |
|
CUY21 Publication List, Electrode Recommendations and Protocol Information |
… |
|
CUY21 Publication List, Electrode Recommendations and Protocol Information |
… |
|
CUY21 Publication List, Electrode Recommendations and Protocol Information |
… |
|
CUY21 Publication List, Electrode Recommendations and Protocol Information |
… |
|
CUY21 Publication List, Electrode Recommendations and Protocol Information |
… |
|
CUY21 Publications by Research Interest |
… |
|
CUY21 Publications by Research Interest |
… |
|
CUY21 Publications by Research Interest |
… |
|
CUY21 Publications by Research Interest |
… |
|
CUY21SC - Square Wave Electroporator |
The CUY21SC measures and displays the delivered current immediately after an EP event. This data is crucial to enable the researcher to verify (on the basis of V=IR) that the correct electroporation protocol has been delivered. As the CUY21SC can measure… |
|
CUY21SC - Square Wave Electroporator |
The CUY21SC measures and displays the delivered current immediately after an EP event. This data is crucial to enable the researcher to verify (on the basis of V=IR) that the correct electroporation protocol has been delivered. As the CUY21SC can measure… |
|
CUY21SC - Square Wave Electroporator |
The CUY21SC measures and displays the delivered current immediately after an EP event. This data is crucial to enable the researcher to verify (on the basis of V=IR) that the correct electroporation protocol has been delivered. As the CUY21SC can measure… |
|
CUY21SC - Square Wave Electroporator |
The CUY21SC measures and displays the delivered current immediately after an EP event. This data is crucial to enable the researcher to verify (on the basis of V=IR) that the correct electroporation protocol has been delivered. As the CUY21SC can measure… |
|
CUY21SC - Square Wave Electroporator |
The CUY21SC measures and displays the delivered current immediately after an EP event. This data is crucial to enable the researcher to verify (on the basis of V=IR) that the correct electroporation protocol has been delivered. As the CUY21SC can measure… |
|
CUY21SC - Square Wave Electroporator |
The CUY21SC measures and displays the delivered current immediately after an EP event. This data is crucial to enable the researcher to verify (on the basis of V=IR) that the correct electroporation protocol has been delivered. As the CUY21SC can measure… |
|
CUY21SC - Square Wave Electroporator |
The resistance of an EP target varies according to physical experimental factors such as the volume of the sample, the buffer used and the distance between the electrodes. Fluctuations in resistance impact on delivered current values and this negatively… |
|
Mouse and Rat Small Tissue Electrode Recommendation and Protocol |
For smaller tissue of less than 0.5mm, we do not recommend the CUY520 series. The reason for this is practical; most clients want to minimize the required volume of DNA as much as possible and do not want to waste DNA. For such applications we believe… |
|
Mouse and Rat Small Tissue Electrode Recommendation and Protocol |
For smaller tissue of less than 0.5mm, we do not recommend the CUY520 series. The reason for this is practical; most clients want to minimize the required volume of DNA as much as possible and do not want to waste DNA. For such applications we believe… |
|
Mouse and Rat Small Tissue Electrode Recommendation and Protocol |
For smaller tissue of less than 0.5mm, we do not recommend the CUY520 series. The reason for this is practical; most clients want to minimize the required volume of DNA as much as possible and do not want to waste DNA. For such applications we believe… |
|
Mouse and Rat Tissue and Brain Slice |
… |
|
Zebrafish Retina Electrode Recommendation |
For Larval Zebrafish Retina (which is quite small, 10microns) it may not be possible to inject DNA buffer.  Accordingly, we suggest that electroporation is performed after the larval zebrafish eye is taken out. Once this is done and one intends… |
|
Zebrafish Retina Electrode Recommendation |
For Larval Zebrafish Retina (which is quite small, 10microns) it may not be possible to inject DNA buffer.  Accordingly, we suggest that electroporation is performed after the larval zebrafish eye is taken out. Once this is done and one intends… |
|
Mouse and Rate Tissue and Brain Slice |
… |
|
Zebrafish Retina Electrode Recommendation |
For Larval Zebrafish Retina (which is quite small, 10microns) it may not be possible to inject DNA buffer.  Accordingly, we suggest that electroporation is performed after the larval zebrafish eye is taken out. Once this is done and one intends… |
|
CUY21SC - Square Wave Electroporator |
Because in vivo impedance (electrical resistance) is generally less stable than in vitro impedance, an accurate measurement of the target sample resistance prior to the electroporation event is the most important critical success factor for accurate… |
|
CUY21SC - Square Wave Electroporator |
The resistance of an EP target varies according to physical experimental factors such as the volume of the sample, the buffer used and the distance between the electrodes. Fluctuations in resistance impact on delivered current values and this negatively… |
|
CUY21SC - Square Wave Electroporator |
Because in vivo impedance (electrical resistance) is generally less stable than in vitro impedance, an accurate measurement of the target sample resistance prior to the electroporation event is the most important critical success factor for accurate… |
|
CUY21SC - Square Wave Electroporator |
Because in vivo impedance (electrical resistance) is generally less stable than in vitro impedance, an accurate measurement of the target sample resistance prior to the electroporation event is the most important critical success factor for accurate… |
|
CUY21SC - Square Wave Electroporator |
Because in vivo impedance (electrical resistance) is generally less stable than in vitro impedance, an accurate measurement of the target sample resistance prior to the electroporation event is the most important critical success factor for accurate… |
|
CUY21SC - Square Wave Electroporator |
Because in vivo impedance (electrical resistance) is generally less stable than in vitro impedance, an accurate measurement of the target sample resistance prior to the electroporation event is the most important critical success factor for accurate… |
|
CUY21SC - Square Wave Electroporator |
Because in vivo impedance (electrical resistance) is generally less stable than in vitro impedance, an accurate measurement of the target sample resistance prior to the electroporation event is the most important critical success factor for accurate… |
|
CUY21SC - Square Wave Electroporator |
Because in vivo impedance (electrical resistance) is generally less stable than in vitro impedance, an accurate measurement of the target sample resistance prior to the electroporation event is the most important critical success factor for accurate… |
|
CUY21SC - Square Wave Electroporator |
Because in vivo impedance (electrical resistance) is generally less stable than in vitro impedance, an accurate measurement of the target sample resistance prior to the electroporation event is the most important critical success factor for accurate… |
|
CUY21SC - Square Wave Electroporator |
Because in vivo impedance (electrical resistance) is generally less stable than in vitro impedance, an accurate measurement of the target sample resistance prior to the electroporation event is the most important critical success factor for accurate… |
|
CUY21SC - Square Wave Electroporator |
Because in vivo impedance (electrical resistance) is generally less stable than in vitro impedance, an accurate measurement of the target sample resistance prior to the electroporation event is the most important critical success factor for accurate… |
|
CUY21SC - Square Wave Electroporator |
Because in vivo impedance (electrical resistance) is generally less stable than in vitro impedance, an accurate measurement of the target sample resistance prior to the electroporation event is the most important critical success factor for accurate… |
|
CUY21 Publication List, Electrode Recommendations and Protocol Information |
… |
|
CUY21 Publication List, Electrode Recommendations and Protocol Information |
… |
|
CUY21 Publication List, Electrode Recommendations and Protocol Information |
… |
|
CUY21 Publication List, Electrode Recommendations and Protocol Information |
… |
|
CUY21SC - Square Wave Electroporator |
Because in vivo impedance (electrical resistance) is generally less stable than in vitro impedance, an accurate measurement of the target sample resistance prior to the electroporation event is the most important critical success factor for accurate… |
|
CUY21SC - Square Wave Electroporator |
Because in vivo impedance (electrical resistance) is generally less stable than in vitro impedance, an accurate measurement of the target sample resistance prior to the electroporation event is the most important critical success factor for accurate… |
|
CUY21 Publication List, Electrode Recommendations and Protocol Information |
… |
|
NEPA21, CUY21SC and CUY21EDIT Multiple Applications and Electrode Recommendations |
A 2 cm midline incision is then made in the abdominal wall along the linea alba using a set of forceps and scissors. A piece of sterile gauze with a hole cut in the center is placed over the incision, and one uterine horn is drawn out through the… |
|
NEP21, CUY21SC and CUY21EDIT Multiple Applications and Electrode Recommendations |
… |
|
NEPA21, CUY21SC and CUY21EDIT Multiple Applications and Electrode Recommendations |
A 2 cm midline incision is then made in the abdominal wall along the linea alba using a set of forceps and scissors. A piece of sterile gauze with a hole cut in the center is placed over the incision, and one uterine horn is drawn out through the hole… |
|
NEPA21_Retina_EP |
NEPA21 Retina EP
Ex Vivo EP
Â
Â
|
|
NEPA21_Retina_EP |
(B) The current was applied with the positive electrode contralateral to the injected eye. After prior injection of plasmid DNA into the subretinal space of the right eye, this arrangement electrophoresed the negatively-charged DNA toward the RPE layer… |
|
NEPA21_Retina_EP |
NEPA21 Retina EP
Ex Vivo EP
Â
Â
|
|
NEPA21_Retina_EP |
… |
|
NEPA21_Retina_EP |
… |
|
NEPA21_Retina_EP |
… |
|
NEPA21_Retina_EP |
(B) The current was applied with the positive electrode contralateral to the injected eye. After prior injection of plasmid DNA into the subretinal space of the right eye, this arrangement electrophoresed the negatively-charged DNA toward the RPE layer… |
|
NEPA21_Retina_EP |
 Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEP21, CUY21SC and CUY21EDIT Multiple Applications and Electrode Recommendations |
… |
|
NEPA21, CUY21SC and CUY21EDIT Multiple Applications and Electrode Recommendations |
… |
|
NEPA21, CUY21SC and CUY21EDIT Multiple Applications and Electrode Recommendations |
A 2 cm midline incision is then made in the abdominal wall along the linea alba using a set of forceps and scissors. A piece of sterile gauze with a hole cut in the center is placed over the incision, and one uterine horn is drawn out through the… |
|
Electroporation Protocols - In Vitro, In Ovo, In Utero, In Vivo, Ex Vivo and New Culture |
… |
|
In Ovo Electroporation |
Some clients, however, who have to contend with a large number of samples on a daily basis find it difficult to manipulate a pair of electrodes vertically. Such clients prefer to move the pair of electrodes horizontally. The CUY613 series was designed… |
|
Zebrafish_Electroporation |
Please note the Zebrafish video referred to above demonstrates the NEPA21 technique used in the article: ‘Direct Delivery of MIF Morpholinos Into the Zebrafish Otocyst by Injection and Electroporation Affects Inner Ear Development (Katie E. Holmes1… |
|
Zebrafish_Electroporation |
Please note the Zebrafish video referred to above demonstrates the NEPA21 technique used in the article: ‘Direct Delivery of MIF Morpholinos Into the Zebrafish Otocyst by Injection and Electroporation Affects Inner Ear Development (Katie E. Holmes1… |
|
Cultured Mouse Embryos and Large Tissue |
For smaller tissue of less than 0.5mm, we do not recommend the CUY520 series. The reason for this is practical; most clients want to minimize the required volume of DNA as much as possible and do not want to waste DNA. For such applications we believe… |
|
Cultured Mouse Embryos and Large Tissue |
For smaller tissue of less than 0.5mm, we do not recommend the CUY520 series. The reason for this is practical; most clients want to minimize the required volume of DNA as much as possible and do not want to waste DNA. For such applications we believe… |
|
Mouse and Rate Tissue and Brain Slice |
… |
|
Cultured Mouse Embryos and Large Tissue |
For smaller tissue of less than 0.5mm, we do not recommend the CUY520 series. The reason for this is practical; most clients want to minimize the required volume of DNA as much as possible and do not want to waste DNA. For such applications we believe… |
|
Mouse and Rate Tissue and Brain Slice |
… |
|
Mouse and Rate Tissue and Brain Slice |
… |
|
Mouse and Rate Tissue and Brain Slice |
… |
|
Mouse and Rat Tissue and Brain Slice |
… |
|
Mouse and Rat Tissue and Brain Slice |
… |
|
NEPA21_Retina_EP |
 Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
 Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
 Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
Mouse and Rat Tissue and Brain Slice |
… |
|
Mouse and Rat Tissue and Brain Slice |
… |
|
Mouse and Rat Tissue and Brain Slice |
… |
|
Mouse and Rat Tissue and Brain Slice |
… |
|
Mouse and Rat Tissue and Brain Slice |
… |
|
Mouse and Rat Tissue and Brain Slice |
… |
|
Mouse and Rat Tissue and Brain Slice |
… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
Zebrafish_Electroporation |
Please note the Zebrafish video referred to above demonstrates the NEPA21 technique used in the article: ‘Direct Delivery of MIF Morpholinos Into the Zebrafish Otocyst by Injection and Electroporation Affects Inner Ear Development (Katie E. Holmes1… |
|
Zebrafish_Electroporation |
Please note the Zebrafish video referred to above demonstrates the NEPA21 technique used in the article: ‘Direct Delivery of MIF Morpholinos Into the Zebrafish Otocyst by Injection and Electroporation Affects Inner Ear Development (Katie E. Holmes1… |
|
Zebrafish_Electroporation |
Please note the Zebrafish video referred to above demonstrates the NEPA21 technique used in the article: ‘Direct Delivery of MIF Morpholinos Into the Zebrafish Otocyst by Injection and Electroporation Affects Inner Ear Development (Katie E. Holmes1… |
|
Zebrafish_Electroporation |
Please note the Zebrafish video referred to above demonstrates the NEPA21 technique used in the article: ‘Direct Delivery of MIF Morpholinos Into the Zebrafish Otocyst by Injection and Electroporation Affects Inner Ear Development (Katie E. Holmes1… |
|
Zebrafish_Electroporation |
Please note the Zebrafish video referred to above demonstrates the NEPA21 technique used in the article: ‘Direct Delivery of MIF Morpholinos Into the Zebrafish Otocyst by Injection and Electroporation Affects Inner Ear Development (Katie E. Holmes1… |
|
Zebrafish_Electroporation |
Please note the Zebrafish video referred to above demonstrates the NEPA21 technique used in the article: ‘Direct Delivery of MIF Morpholinos Into the Zebrafish Otocyst by Injection and Electroporation Affects Inner Ear Development (Katie E. Holmes1… |
|
In Ovo Electroporation |
Some clients, however, who have to contend with a large number of samples on a daily basis find it difficult to manipulate a pair of electrodes vertically. Such clients prefer to move the pair of electrodes horizontally. The CUY613 series was designed… |
|
In Ovo Electroporation |
Some clients, however, who have to contend with a large number of samples on a daily basis find it difficult to manipulate a pair of electrodes vertically. Such clients prefer to move the pair of electrodes horizontally. The CUY613 series was designed… |
|
Cultured Mouse Embryos and Large Tissue |
For smaller tissue of less than 0.5mm, we do not recommend the CUY520 series. The reason for this is practical; most clients want to minimize the required volume of DNA as much as possible and do not want to waste DNA. For such applications we believe… |
|
Cultured Mouse Embryos and Large Tissue |
For smaller tissue of less than 0.5mm, we do not recommend the CUY520 series. The reason for this is practical; most clients want to minimize the required volume of DNA as much as possible and do not want to waste DNA. For such applications we believe… |
|
Cultured Mouse Embryos and Large Tissue |
For smaller tissue of less than 0.5mm, we do not recommend the CUY520 series. The reason for this is practical; most clients want to minimize the required volume of DNA as much as possible and do not want to waste DNA. For such applications we believe… |
|
Mouse and Rat Tissue and Brain Slice |
… |
|
Mouse and Rat Tissue and Brain Slice |
… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
Mouse and Rat Tissue and Brain Slice |
… |
|
Zebrafish_Electroporation |
Please note the Zebrafish video referred to above demonstrates the NEPA21 technique used in the article: ‘Direct Delivery of MIF Morpholinos Into the Zebrafish Otocyst by Injection and Electroporation Affects Inner Ear Development (Katie E. Holmes1… |
|
Zebrafish_Electroporation |
Please note the Zebrafish video referred to above demonstrates the NEPA21 technique used in the article: ‘Direct Delivery of MIF Morpholinos Into the Zebrafish Otocyst by Injection and Electroporation Affects Inner Ear Development (Katie E. Holmes1… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
Arrangement of electrodes for in vivo electroporation for RPE transfection.(A) Tweezer-type electrodes were placed on the corneal surface of either eye of a 1-month-old Sprague-Dawley rat.(B) The current was applied with the positive electrode contralateral… |
|
NEPA21_Retina_EP |
(B) The current was applied with the positive electrode contralateral to the injected eye. After prior injection of plasmid DNA into the subretinal space of the right eye, this arrangement electrophoresed the negatively-charged DNA toward the RPE layer… |
|
NEPA21_Retina_EP |
(B) The current was applied with the positive electrode contralateral to the injected eye. After prior injection of plasmid DNA into the subretinal space of the right eye, this arrangement electrophoresed the negatively-charged DNA toward the RPE layer… |
|
NEPA21_Retina_EP |
(B) The current was applied with the positive electrode contralateral to the injected eye. After prior injection of plasmid DNA into the subretinal space of the right eye, this arrangement electrophoresed the negatively-charged DNA toward the RPE layer… |
|
NEPA21_Retina_EP |
(B) The current was applied with the positive electrode contralateral to the injected eye. After prior injection of plasmid DNA into the subretinal space of the right eye, this arrangement electrophoresed the negatively-charged DNA toward the RPE layer… |
|
NEPA21_Retina_EP |
(B) The current was applied with the positive electrode contralateral to the injected eye. After prior injection of plasmid DNA into the subretinal space of the right eye, this arrangement electrophoresed the negatively-charged DNA toward the RPE layer… |
|